Main Characters
|
Description
|
Measurement object
|
Oxygen concentration in gaseous mixture
|
Measurement system
|
Paramagnetic system
|
Measurement range
|
0-5 to 0-25 vol%O2, 3 range can be programmed
|
Self-diagnostic content
|
Sensor unit error, Constant temperature chamber error, Analog error, Memory error, Calibration coefficient error
|
Analog output signal
|
4 to 20 mA DC (load resistance; Max. 550Ω)
|
Contact output
|
Fail; 1 point, open or closed when error occurs, user configurable, Maintenance status; 1 point, closed during maintenance, Range answerback or Hi/Lo alarm; 2 points, normally de-energized
|
Contact input
|
Remote range switching, calibration start
|
Calibration methods
|
Automatic, semiautomatic and manual calibration
|
Zero gas
|
Nitrogen, not contain O2 gas with a concentration equal to or greater than 0.1% of the upper range value,
|
Span gas
|
Dry air, instrument air or standard gas containing O2 gas with a concentration of 80 to 100 %
|
Auxiliary gas pressure
|
N2 gas, 180 kPa(Approx. 35 ml/min)
|
Sample gas condition
|
Gas flow; 200 ml/min ±10%, Gas temperature; 0 to 50℃, Humidity; No moisture condensation in the flow path or the sensor
|
Ambient temperature
|
5 to 55℃
|
Installation
|
Indoor, panel or wall mounting
|
Power supply
|
100 to 115 V AC or 200 to 240 V AC
|